• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CYBERONICS, INC. PULSE GEN MODEL 106; GENERATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CYBERONICS, INC. PULSE GEN MODEL 106; GENERATOR Back to Search Results
Model Number 106
Device Problems No Device Output (1435); Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/02/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that the physician had a patient who he could not get results from the heart rate detection feature.Some initial troubleshooting steps were performed, including attempting sensitivity settings 4 and 5, but this did not resolve the issue.The patient did not have pre-surgical evaluation done at the time of implant.The generator dhr was reviewed and found that all specifications were met prior to distribution.No additional pertinent information has been received to date.
 
Event Description
It was later reported on (b)(6) 2016 that the patient's vns system was recently interfaced with to troubleshoot the heart rate verification feature.This was after the previous reported attempt on (b)(6) 2016.The output currents were disabled on the generator.The heart rate verification was attempted and showed a stable heart rate value.No issues of difficulties were encountered while registering the heart rate.The generator's output currents were turned back on after the testing.No changes were made to the sensitivity setting throughout the testing.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PULSE GEN MODEL 106
Type of Device
GENERATOR
Manufacturer (Section D)
CYBERONICS, INC.
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
CYBERONICS, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key5752705
MDR Text Key49031332
Report Number1644487-2016-01434
Device Sequence Number1
Product Code LYJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 06/02/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date09/03/2017
Device Model Number106
Device Lot Number203513
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Event Location Other
Initial Date Manufacturer Received 06/02/2016
Initial Date FDA Received06/27/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received09/14/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/16/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age58 YR
-
-