• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CONMED CORPORATION 1/4 INCH X 6 FOOT SUCTION TUBING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CONMED CORPORATION 1/4 INCH X 6 FOOT SUCTION TUBING Back to Search Results
Catalog Number 0037860
Health Effect - Clinical Code No Patient Involvement (2645)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Conmed corporation received four (4) "unopened" packages containing the 1/4" x 6' suction tubing for evaluation.Visual examination of the returned packages found all four (4) packages with partial seal disruptions, all occurring in the final manufacturing seals.In all four (4) of the returned packages, there is evidence that the device, suction tubing, was sealed within the sterile seal of the package.In those areas adhesive transfer to the poly material showing that a full seal existed is missing due to the product obstructing the seal.The devices were transferred to packaging engineering for evaluation and one (1) device was confirmed as unsterile failing the dye leak testing on 03-jun-2016.The other three (3) devices were confirmed for intact sterile barrier.This lot was manufactured on 08-jun-2015.A review of the device history record for this lot found no ncrs and no note of discrepancies during the manufacturing process.In these instances, preliminary investigation revealed that the most probable cause of the seal disruptions appeared to be related to the device being within the sterile seal area during the final manufacturing sealing process.(b)(4).The reported packaging anomalies were obvious to the distributor and therefore prompted the return of the devices for evaluation and replacement.To date, there have been no serious injuries or death related to this reported problem.Nonetheless, to prevent future recurrences, an investigation has been opened to address this issue.This information has been forwarded to the investigation owner.As with all medical devices, examination of packaging occurs multiple times prior to use (shipping/receiving, distribution, storage and prior to use).In addition, good clinical practice would include examination and verification of the product and its original packaging to ensure both are intact.If the product and its packaging have been opened/damaged or altered, do not use the product and contact the manufacturer immediately.
 
Event or Problem Description
The distributor in (b)(4) reported that during receiving and inspection of incoming products, four (4) packages containing 1/4" x 6' suction tubing were discovered with "insufficient heat seal".All four (4) of the returned packages exhibited partial seal disruption in the final manufacturing seals.In this instance, there was no patient involvement with this reported problem, as the packaging anomaly was discovered during inspection at the distributor facility prior to distribution to an end-user.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
1/4 INCH X 6 FOOT SUCTION TUBING
Common Device Name
SUCTION TUBING
Manufacturer (Section D)
CONMED CORPORATION
525 french road
utica NY 13502 5994
MDR Report Key5756079
Report Number3007305485-2016-00071
Device Sequence Number185803
Product Code BYY
UDI-Device Identifier20653405000621
UDI-Public(01)20653405000621(17)200606(30)50(10)201506085
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
P/A EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,foreign,other
Initial Reporter Occupation Other
Remedial Action Other
Type of Report Initial
Report Date (Section B) 06/03/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Expiration Date06/06/2020
Device Catalogue Number0037860
Device Lot Number201506085
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/27/2016
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/03/2016
Initial Report FDA Received Date06/28/2016
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-