STRYKER ORTHOPAEDICS-MAHWAH SIZE 4-8 KEEL PUNCH GUIDE; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO
|
Back to Search Results |
|
Catalog Number 6541-2-748 |
Device Problems
Crack (1135); Degraded (1153); Peeled/Delaminated (1454); Metal Shedding Debris (1804); Material Integrity Problem (2978)
|
Patient Problems
No Patient Involvement (2645); No Known Impact Or Consequence To Patient (2692)
|
Event Date 06/01/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
When completed, the investigation results will be submitted in a supplemental report.
|
|
Event Description
|
Rubber handles are old, peeling, cracked and shedding.Update: issue noticed during sterile procedure or while instrument was cleaning.Not during the regular procedure.
|
|
Manufacturer Narrative
|
An event regarding santoprene damage involving a triathlon keel punch guide was reported.A review of the device history records indicates that the reported devices were manufactured and accepted into final stock with no reported discrepancies.The complaint history review indicated that there were no similar events for the reported lot.Conclusion: visual inspection showed the device handle has discolored and damaged.The damaged device was discovered during inspection.There was no surgical procedure associated with the reported event.This event meets the definition of preventive maintenance.No further investigation is required at this time.If additional information becomes available to indicate further evaluation is warranted, this record will be reopened.
|
|
Event Description
|
Rubber handles are old, peeling, cracked and shedding.Update: issue noticed during sterile procedure or while instrument was cleaning.Not during the regular procedure.
|
|