• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC - GALWAY SYNERGY¿; CORONARY DRUG-ELUTING STENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - GALWAY SYNERGY¿; CORONARY DRUG-ELUTING STENT Back to Search Results
Model Number H7493926220250
Device Problems Device Damaged Prior to Use (2284); Material Deformation (2976)
Patient Problem No Patient Involvement (2645)
Event Date 03/08/2016
Event Type  malfunction  
Manufacturer Narrative
Device is a combination product.(b)(4).Device evaluated by mfr: the stent delivery system (sds) was returned for analysis.There was no issue with the hub.There were several stent struts that were stretched and bent.The stent had stretched over the distal cone of the balloon.The bumper tip of the device showed no signs of damage.The balloon cone profiles were reviewed and no issues were noted with the overall balloon profile.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.A visual and tactile examination found no issues with the hypotube shaft profile or the shaft polymer extrusion profile.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The root cause is not confirmed as there was no evidence of the alleged issue or any anomalies which could have contributed to the reported difficulty.(b)(4).
 
Event Description
Reportable based on device analysis completed on 01-jun-2016.It was reported that hub damage occurred.A 2.50x20 synergy¿ dug-eluting stent was selected for use.During unpacking, it was noted that the hub of the stent visually seemed to be slightly crushed.No patient complications were reported.However, returned device analysis revealed stent damage.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNERGY¿
Type of Device
CORONARY DRUG-ELUTING STENT
Manufacturer (Section D)
BOSTON SCIENTIFIC - GALWAY
Manufacturer (Section G)
BOSTON SCIENTIFIC - GALWAY
Manufacturer Contact
linda leimer
one scimed place
maple grove, MN 55311
7634941700
MDR Report Key5759468
MDR Text Key48467139
Report Number2134265-2016-05460
Device Sequence Number1
Product Code NIQ
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
SIMILAR
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Health Professional
Type of Report Initial
Report Date 06/01/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/25/2017
Device Model NumberH7493926220250
Device Catalogue Number39262-2025
Device Lot Number18275989
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/08/2016
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/01/2016
Initial Date FDA Received06/29/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/19/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-