DEPUY SYNTHES POWER TOOLS MINI AIR DRILL; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
|
Back to Search Results |
|
Catalog Number 512.100 |
Device Problems
Defective Component (2292); Failure to Shut Off (2939)
|
Patient Problem
No Patient Involvement (2645)
|
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(6).The device manufacture date is currently not available.This device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.Reliability engineering evaluated the device and the reported condition was confirmed.It was further determined that the motor was not functioning.The assignable root cause was determined to be due to improper maintenance, which is user error, misuse, and / or abuse.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|
Event Description
|
It was reported by (b)(6) that during service and evaluation, it was discovered that the motor was defective on the mini air drill device.It was further determined that after connecting the device, the drill did not stop.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of this event was unknown.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|
Manufacturer Narrative
|
In addition to the malfunctions identified in the initial report, (b)(4) also determined that the device did not work.It was further determined that the device was very dry inside allowing the engine/motor to stall.If information is obtained that was not available, a follow-up medwatch will be filed as appropriate.
|
|
Search Alerts/Recalls
|
|
|