| Catalog Number |
03822601 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Codes |
Injury (2348); No Code Available (3191)
|
| Date of Event |
06/07/2016
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that; the second revision surgery was performed due to the rod breakage.The left and right of the rod were broke in two levels (left: l4/5 and l3/4, right: l5/s1 and l4/5).During the revision surgery, the surgeon tried to remove the closed polyaxial screw on left s2ai.But the screw was not able to removed and the tip of the drivers was deformed.The screw shaft was remained.This event is related to (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Method: risk assessment; result: no lot # was provided, so a manufacturing record review could not be performed.The product is not returned for evaluation and material analysis.The surgical technique for xia states " do not exceed 12nm during final tightening." excessive force applied and over tightening screws during final tightening could lead to excessive stress on the construct which could lead to stress concentration leading to rod fracture post- op.Conclusion: the definitive cause of the event could not be determined with the given information.
|
| |
|
Event or Problem Description
|
|
It was reported that; the second revision surgery was performed due to the rod breakage.The left and right of the rod were broke in two levels (left : l4/5 and l3/4, right : l5/s1 and l4/5).During the revision surgery, the surgeon tried to remove the closed polyaxial screw on left s2ai.But the screw was not able to removed and the tip of the drivers was deformed.The screw shaft was remained.This event is related to (b)(4).
|
| |
|
Search Alerts/Recalls
|