Catalog Number QDP-20X5.5
Medical Device Problem Codes Hole In Material (1293); Use of Device Problem (1670)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 06/10/2016
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Concomitant product: boston scientific inflation handle, cook syringe (unknown type).Investigation evaluation: a product evaluation was not performed in response to this report because the product said to be involved was not provided to cook for evaluation.The report could not be confirmed.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusions: we could not conduct a complete investigation because the product said to be involved was not returned for evaluation.A definitive cause for the reported observation could not be determined.The additional information provided indicated the balloon did not receive lubrication prior to advancement through the endoscope.This is the most likely cause for the reported observation.The instructions for use direct the user to "apply a water soluble lubricant to balloon to allow easier passage through accessory channel." this activity will aid in endoscopic advancement and balloon preservation.Another possible contributing factor to balloon damage is failure to apply negative pressure to the balloon dilator prior to advancement through the endoscope.The instructions for use direct the user, "prior to insertion of balloon dilator, negative pressure is mandatory to maintain balloon deflation." the application of negative pressure will aspirate all residual air from the balloon and ease endoscopic advancement.Negative pressure will also aid in balloon preservation and optimize balloon performance.Damage to the balloon material can occur if the balloon is inflated prior to advancement through the endoscope or if the balloon is inflated while partially or fully inside the accessory channel of the endoscope.The instructions for use contain the following precaution: ¿do not pre-inflate balloon.¿ prior to distribution, all quantum ttc pyloric balloon dilators are subjected to a visual examination to ensure device integrity.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: a review of the complaint history was conducted.The likelihood of occurrence is considered rare.Corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.Additional comments regarding this report: based on the information provided, a cook representative has been directed to contact the medical facility involved in an effort to promote further education and understanding related to appropriate usage of this product.
 
Event or Problem Description
During an esophagogastroduodenoscopy (egd), the physician used a cook quantum ttc pyloric balloon dilator.The quantum ttc pyloric balloon was opened [and taken] out of the pack and reverse suction was used using an alliance gun & cook syringe.The balloon was entered down the endoscope and a hole was noted when attempting to inflate.Another device was used to complete the procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUANTUM TTC PYLORIC BALLOON DILATOR
Common Device Name
KNQ, DILATOR, ESOPHAGEAL
Manufacturer (Section D)
COOK ENDOSCOPY
4900 bethania station rd
winston-salem NC 27105
MDR Report Key5768877
Report Number1037905-2016-00194
Device Sequence Number647547
Product Code KNQ
UDI-Device Identifier00827002229983
UDI-Public(01)00827002229983(17)190228(10)W3687816
Combination Product (Y/N)N
Initial Reporter StateSC
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 06/10/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberQDP-20X5.5
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Distributor Facility Aware Date06/10/2016
Device Age4 MO
Event Location Hospital
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/10/2016
Initial Report FDA Received Date07/05/2016
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-