Catalog Number IAP-0500F |
Device Problem
Power Problem (3010)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 04/15/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).
|
|
Event Description
|
It has been reported via a field service report: (b)(4).Symptom: pump turns off in use findings / actions taken: power supply replaced.Pump checked and pass.Software: (b)(4).Reason for service: other reason update on 06/22/2016-per the field service agent "there were no reports of complaints.".
|
|
Manufacturer Narrative
|
Qn#(b)(4).Additional information received 07/18/2016 the pump was on a patient at the time of the event.Pump switched.The condor power supply (p/n: 77-0063-001 s/n: (b)(4)) was returned for evaluation.Visual inspection of condor power supply was performed and found no obvious damage or defects.The condor power supply was installed into a known good autocat2w and functional testing was performed.The pump was startup as normal.A power supply voltage check was performed per autocat2 series service manual (rev 3) and all voltages were within specification.The pump was run on battery (battery load test) until the iabp shut down (>90minutes within spec).A battery load test was performed after letting the pump charge for over 8 hours and the unit ran for over 90 minutes.The condor power supply passed all functional testing.Visual inspection of the condor power supply internal hardware was performed and no component damage or defect was noted.See other remarks section.Other remarks: a device history record (dhr) review was conducted for the lot number/serial number with no relevant findings.The device passed all manufacturing specifications prior to release.Conclusion: the reported complaint of "turns off in use" is not confirmed.The reported problem could not be reproduced during the functional test.The power supply passed functional testing.The cause of the reported complaint is undetermined.
|
|
Event Description
|
It has been reported via a field service report: (b)(4).Symptom: pump turns off in use.Findings / actions taken: power supply replaced.Pump checked and pass.Software: 2.24fr.Fcn: (b)(4).Reason for service: other reason.Update on 06/22/2016-per the field service agent "there were no reports of complaints.".
|
|
Search Alerts/Recalls
|