Catalog Number IAP-0500 |
Device Problem
Gas/Air Leak (2946)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 06/09/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).
|
|
Event Description
|
It was reported via a field service report - (b)(4).Pump had purge failure alarm when first put on patient.Pump was switched out quickly without delay or impact to patient.Findings/actions taken: confirmed- pcs assembly replaced.Parts sent to and replaced by hospital biomed.Defective pcs assy.To be returned on (b)(4).Fcn level: 1416, software level: 2.24.
|
|
Manufacturer Narrative
|
Qn#(b)(4).The pcs assembly (p/n 96-3006-001, s/n (b)(4)) was returned for evaluation.Visual inspection of the pcs assembly was performed and no abnormality was found.The pcs assembly in question was installed into known good autocat2w and performed functional testing.The known good autocat2w with the pcs assembly in question installed passed the functional test.The purge cycle was performed multiple times with no alarms or errors.The pump was then left to run for over six hours without any issues.The pcs assembly in question was installed onto the mdt-50 leak tester and passed leak testing.Visual inspection of the pcs assembly internal hardware was performed and no abnormality was found.Per service history reported, the pcs assembly was replaced on november 22, 2005.A device history record (dhr) review was conducted for the iabp serial/ lot number with no relevant findings.The device passed all manufacturing specifications prior to release.Other remarks: conclusion: the reported problem of purge failure alarm is not confirmed.The reported problem could not be reproduced during the functional test.The pcs assembly passed functional testing.The cause of the reported complaint is undetermined.
|
|
Event Description
|
It was reported via a field service report - l610990.Pump had purge failure alarm when first put on patient.Pump was switched out quickly without delay or impact to patient.Findings/actions taken: confirmed- pcs assembly replaced.Parts sent to and replaced by hospital biomed.Defective pcs assy.To be returned on (b)(4).Fcn level: 1416, software level: 2.24.
|
|
Search Alerts/Recalls
|