| Catalog Number |
48811339 |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
06/14/2016
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that; the locking wing on plate bent upon screw insertion due to wing overhanging the screw hole.
|
| |
|
Additional Manufacturer Narrative
|
|
Method: risk assesment; result: the plate was confirmed by the stryker rep to have a deformed spring bar.The stryker rep confirmed that no drill guides were used to prepare the screw holes.Conclusion: the cause is not using the recommended drill guides as per the stg.
|
| |
|
Event or Problem Description
|
|
It was reported that; the locking wing on plate bent upon screw insertion due to wing overhanging the screw hole.
|
| |
|