Model Number MERGE HEMODYNAMICS 10.0.1
Device Problems Device Emits Odor (1425); Overheating of Device (1437)
Patient Problem No Patient Involvement (2645)
Event Date 06/10/2016
Event Type  malfunction  
Manufacturer Narrative
The faulty ups was returned by the customer to merge healthcare on june 21, 2016 for evaluation and testing.Results showed that the batteries in the unit were defective.The batteries were replaced and recharged.The unit then underwent qc testing.Upon successful completion of the testing, the unit was returned to service stock.
 
Event Description
Merge hemodynamics monitors, measures, and records physiological data from a human patient undergoing a cardiac catheterization procedure.The system comprises the patient data module and the merge hemodynamics hemo monitor pc.The two units are connected via a serial interface.All vital parameters and evaluations are registered and calculated in the patient data module.This data is then transmitted to the merge hemodynamics hemo monitor pc via the serial interface.All data can be shown and monitored on the merge hemodynamics hemo monitor pc.On (b)(6) 2016, a customer reported to merge healthcare that the ups (uninterrupted power supply), a sealed battery containing lead and sulfuric acid, emitted a burning odor and caused the hemo computers to shut down.The customer confirmed that no patient was present when the problem occurred.If parts of the hemo system become energized, there is a potential for direct harm to the patient and/or user including electrical shock or burns.However, the customer confirmed that no one was neither injured nor needed treatment as a result of this event.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MERGE HEMODYNAMICS
Type of Device
PROGRAMMABLE DIAGNOSTIC COMPUTER
Manufacturer (Section D)
MERGE HEALTHCARE
900 walnut ridge drive
hartland WI 53029
Manufacturer (Section G)
MERGE HEALTHCARE
900 walnut ridge drive
hartland WI 53029
Manufacturer Contact
michael diedrick
900 walnut ridge drive
hartland, WI 53029
2629123570
MDR Report Key5781440
MDR Text Key49127926
Report Number2183926-2016-00633
Device Sequence Number1
Product Code DQK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K082421
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial
Report Date 06/10/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/10/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMERGE HEMODYNAMICS 10.0.1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/21/2016
Is the Reporter a Health Professional? No
Date Manufacturer Received06/10/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/12/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-