| Model Number |
179094F |
| Medical Device Problem Codes |
Dull, Blunt (2407); Failure to Cut (2587)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
01/27/2016
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Manufacturer reference number: (b)(4).Since the lot number was not provided, this information cannot be determined.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during a donor nephrectomy, at the first usage there were difficulties while entering to the abdomen and the device could not cut.There is no problem related cannula.Knife/cutter did not cut.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Post market vigilance (pmv) and concurrently with engineering led an evaluation of five trocars with fixation cannula opened by the account.Visual inspection noted the five obturators were received heat damaged.The tops of the obturators were not seated properly.A trocar was received.Functionally, the condition of the instrument precludes functional testing.Replication of the observed heat damage may occur if the instrument is subjected to excessive heat during a sterilization process.Should new information become available, the file will be re-opened and the investigation summary amended as appropriate.
|
| |
|