Model Number MICL12.6
Medical Device Problem Codes Inadequacy of Device Shape and/or Size (1583); Appropriate Device Problem Term/Code Not Available (3191)
Health Effect - Clinical Codes Headache (1880); Pain (1994); Blurred Vision (2137); Vomiting (2144); No Code Available (3191)
Date of Event 06/07/2016
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Results: visual inspection of the returned product found pieces of one haptic torn off and missing.The lens was returned in a small clear container.(b)(4).
 
Event or Problem Description
The reporter stated the surgeon implanted a 12.6mm micl12.6 implantable collamer lens, -12.0 diopter, in the patient's left eye (os) on (b)(6) 2016.The lens was explanted on (b)(6) 2016 due to the lens being too large and hypotony (low iop).The patient experienced blurred visual acuity, eye pain, headache, and vomiting.The reporter indicated that the cause of the event was unknown.The patient's post-op best-corrected visual acuity (bcva) was 20/25.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ICL (IMPLANTABLE COLLAMER LENS)
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
MDR Report Key5788401
Report Number2023826-2016-00898
Device Sequence Number16485467
Product Code MTA
UDI-Device Identifier00841542103060
UDI-Public(01)00841542103060(17)171231
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
P030016
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2016
Reporter Type Manufacturer
Report Source user facility
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 06/17/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Physician
Device Expiration Date12/31/2017
Device Model NumberMICL12.6
Device Catalogue NumberN/A
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/29/2016
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/17/2016
Initial Report FDA Received Date07/12/2016
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age43 YR
Patient Weight71
-
-