Upon receipt at our post market quality assurance laboratory, this lead was confirmed to be fractured 40 cm from the terminal pin.Based on the analysis results, we suspect that fatigue or stress in the region of the fracture site due to relative motion between the suture sleeve and an anatomical feature led to the fracture.As a lead moves in response to normal heart rhythms and blood flow, extensive flexing over a period of time may cause fatigue or stress, weakening the coil, ultimately resulting in a fracture.This can occur between the suture sleeve and some other part of the anatomy.Fatigue fractures in the pocket or clavicular/first rib area are well known and documented in the industry.A combination of lead design, implant techniques, and patient anatomy and activity level contribute to these types of occurrences.
|
|
Boston scientific received information that this left ventricular (lv) lead exhibited low sensed amplitudes, high pacing thresholds, high out-of-range pace impedance measurements, and was suspected of a fracture.The patient reported that he had felt "something pop" shortly after implant of this lead when he had caught a heavy item.Subsequently, this lv lead was explanted and replaced.No additional adverse patient effects were reported.
|