Model Number T165
Device Problems Over-Sensing (1438); Low impedance (2285); Ambient Noise Problem (2877)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/09/2016
Event Type  Injury  
Manufacturer Narrative
As no further information concerning this report is expected, our investigation is complete.This investigation will be updated should further information be provided.
 
Event Description
Boston scientific received information that this implantable cardioverter defibrillator (icd) and right ventricular (rv) lead exhibited low out-of-range pace impedance measurements in addition to episodes of noise and oversensing stored to device memory.Neither the low impedance measurements or noise and oversensing could be reproduced when tested in clinic.Despite this, a lead insulation issue was suspected but could not be confirmed via x-ray or visually during revision at which time the rv lead was surgically abandoned and icd explanted.A new device and lead were then implanted without consequence.It was noted that the explanted device is not available for return and analysis.No additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VITALITY 2
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer (Section G)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer Contact
sonali vasekar
4100 hamline ave. n
st. paul, MN 
6515824786
MDR Report Key5789614
MDR Text Key49373593
Report Number2124215-2016-08694
Device Sequence Number1
Product Code KRG
Combination Product (y/n)N
PMA/PMN Number
P960040/S039
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Physician
Type of Report Initial
Report Date 05/10/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date10/08/2010
Device Model NumberT165
Other Device ID NumberVITALITY 2 DR
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Initial Date Manufacturer Received 05/10/2016
Initial Date FDA Received07/13/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/13/2009
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
0185
Patient Outcome(s) Hospitalization; Life Threatening; Required Intervention;
Patient Age78 YR
-
-