| Catalog Number |
48335310 |
| Medical Device Problem Codes |
Mechanical Problem (1384); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Codes |
No Known Impact Or Consequence To Patient (2692); No Information (3190)
|
| Date of Event |
06/21/2016
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that; ring on screw came off during insertion.Drill guides were not used-screws were free-handed.
|
| |
|
Additional Manufacturer Narrative
|
|
Causes of the event are freehand use during screw insertion, inserter loosening during surgery, inserting not fully threading, user unfamiliar with system, interaction between cage/screw/inserter, outer diameter of clip is too large.The established cause of the event is general use error.
|
| |
|
Event or Problem Description
|
|
It was reported that; ring on screw came off during insertion.Drill guides were not used-screws were free-handed.
|
| |
|