| Model Number |
4458 |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
No Information (3190)
|
| Date of Event |
04/03/2016
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
This lead has been returned.Boston scientific has concluded it is unlikely that lead characteristics caused or contributed to the reported clinical observations.As no further information concerning this report is expected, our investigation is complete.This investigation will be updated should further information be provided.
|
| |
|
Event or Problem Description
|
|
Boston scientific received information that this right ventricular (rv) lead was dislodged with tissue in the tines noted.Additional information was obtained indicating that dislodgement was confirmed through x-ray.The rv lead was explanted.No additional adverse patient effects were reported.
|
| |
|