A nurse at a physician's office reported that their handheld programmer was not holding a charge.When the device was plugged in it would start, but then only displayed the windows background.The nurse was instructed to turn off the device, remove the software card and re-insert it, which resolved the issue however, the battery would still not hold a charge.The programmer has not been received by the manufacturer to-date.Additional relevant information has not been received to-date.
|
Analysis was completed on the returned handheld.During the analysis, the main battery was able to hold a charge and power the handheld for over an hour.It was identified that the handheld would not power on.The cause for the anomaly is associated with a swollen main battery.The swollen battery bent the battery cover causing the battery latch switch to register that the latch was unlocked, thus preventing the handheld from powering on.This condition can also cause the handheld to power off intermittently.No further anomalies were identified.No anomalies were identified with the flashcard and the software performed according to functional specifications.
|