INTERVASCULAR SAS INTERGARD KNITTED GRAFT; VASCULAR POLYESTER GRAFT Back to Search Results
Model Number IGK1608
Medical Device Problem Code Manufacturing, Packaging or Shipping Problem (2975)
Health Effect - Clinical Code Blood Loss (2597)
Date of Event 06/24/2016
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A review of the complaint device history records, including collagen coating records, indicated that the graft was processed and inspected according to procedures and no anomaly was found.Specifically, the review of the water permeability testing records of products coated on the same day and under the same conditions as the complaint device indicated values well within product specifications.One retention sample coated on the same period and under the same conditions as the complaint device underwent water permeability testing at 120mmhg as per iso 7198.The test result indicated a value well within product specifications.Remaining fragments of the involved product have been sent to an outside laboratory for macroscopic and scanning electron microscopy analysis.The investigation is still on-going.A follow-up report will be submitted upon completion of the investigation.
 
Event or Problem Description
During an abdominal aortic aneurysm surgery performed on (b)(6) 2016, it was reported a blood leaking from the entire surface of the graft when declamped after the anastomosis of the left common iliac artery.The bleeding stopped by itself after half an hour.The graft remained implanted.No adverse consequence for the patient has been reported.
 
Additional Manufacturer Narrative
(b)(4) remaining fragments of the involved device were analysed by an outside and independant laboratory.It consisted in a macroscopic examination and a scanning electron microscopy analysis.The laboratory concluded that no obvious anomaly was observed at macroscopy.The sem analysis pointed out presence of collagen material infiltration.No tears, filament rupture or signs of sectioning, loss of the textile cohesion, holes were observed.No conclusion can be drawn.However, all available information and the product testing performed would tend to indicate that the device was not defective.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTERGARD KNITTED GRAFT
Common Device Name
VASCULAR POLYESTER GRAFT
Manufacturer (Section D)
INTERVASCULAR SAS
zi athelia 1
la ciotat cedex, 13705
FR  13705
MDR Report Key5805076
Report Number1640201-2016-00018
Device Sequence Number556931
Product Code DSY
UDI-Device Identifier00384401000488
UDI-Public00384401000488
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K964625
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/24/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Expiration Date08/01/2016
Device Model NumberIGK1608
Device Catalogue NumberIGK1608
Device Lot Number11H25
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/04/2016
Type of Report(Section G)FDA Requested
Initial Date Received by Manufacturer 06/24/2016
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date07/19/2016
Supplement Report FDA Received Date11/14/2016
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-