During review of the in-house programming/diagnostic history database, it was observed that on office visit on (b)(6) 2014 the patient's settings were different than what were programmed at the previous office visit on (b)(6) 2014.The settings found were indicative of a faulted diagnostic test; however, review of system diagnostic testing from office visit on (b)(6) 2014 was within normal limits.The device was interrogated prior to the patient leaving the office on (b)(6) 2014 as recommended by device manufacturer to ensure the device is at the correct settings; therefore, the change in settings occurred after the (b)(6) 2014 visit.The physician indicated that the patient was not seen in between these two dates; however, it is unknown if the patient was seen by another physician during this time.No patient adverse events were reported.No additional relevant information has been received to date.
|