| Model Number |
28-95-80 |
| Medical Device Problem Codes |
Failure to Pump (1502); Pumping Stopped (1503)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
06/21/2016
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
There was no patient involvement.Sorin group (b)(4) manufactures the s5 control panel mast roller pump 150.The incident occurred in (b)(6).This medwatch report is filed on behalf of sorin group (b)(4).Sorin group (b)(4) received a report that a pump stop message appeared on the s5 control panel mast roller pump during in-service when using pump #1, and the pump for stop link could not be accessed when pump #1 was selected.There was no patient involvement.The service representative on-site tested the unit and found that the pump worked when set to a different pump number.The mast roller pump control panel was replaced.The investigation is ongoing.A follow-up report will be sent when the investigation is complete.
|
| |
|
Event or Problem Description
|
|
Sorin group (b)(4) received a report that a pump stop message appeared on the s5 control panel mast roller pump during in-service when using pump #1, and the pump for stop link could not be accessed when pump #1 was selected.There was no patient involvement.
|
| |
|