The pipeline device has not been returned for evaluation.The reported event could not be confirmed.The cause of the event could not be conclusively determined from the reported information.Mdrs related to this event: 2029214-2016-00597, 2029214-2016-00598, 2029214-2016-00599.
|
The pipeline and microcatheter were returned for evaluation.The pipeline braid was observed to be partially deployed from the catheter tip for 1.5cm and released from the capture coil.For further examination, the pipeline pushwire was pushed out and the pipeline braid was deployed out of the microcatheter with no issues.The capture coil was observed to be compressed.The pipeline braid was observed to be fully open, with both ends having slight fraying.Based on the analysis findings the clinical observation could not be confirmed.The cause for the reported experience could not be determined as the returned pipeline braid was fully opened, and observed to be released from the capture coil.It is possible that the damaged braid (fraying) may have contributed to the reported issue.In addition, damage was found on the pushwire (bending).However, the cause for the damages could not be determined.All devices are 100% inspected for damage and irregularities during the manufacturing process.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|