ANGIODYNAMICS, INC. DURAFLOW 2 CHRONIC HEMODIALYSIS CATHETER Back to Search Results
Catalog Number H787103012155
Medical Device Problem Code Detachment Of Device Component (1104)
Health Effect - Clinical Code Device Embedded In Tissue or Plaque (3165)
Date of Event 06/22/2016
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The reported defective sheath introducer has been returned to the manufacturer along with a scar, and an investigation into the root cause for event is currently in progress.The results of the device evaluation will be sent via a follow up medwatch.A review of the device history records was performed for the packaging and component lots for any deviations related to the reported defect of the complaint.The review confirms that the lots met all material, assembly, and performance specifications.(b)(4).
 
Event or Problem Description
As reported (b)(6) 2016, a patient of unknown age and gender presented for a catheter placement.During the procedure, while the treating physician was placing the dialysis catheter in the patient's groin, the handles of the valved sheath introducer broke off.As a result of the fracture, 2 pieces of the sheath were left in patient's vein in the groin.The fractured pieces were removed, and the procedure was successfully completed.There was no report of permanent harm or injury to the patient due to this event.The reported defective disposable device has been returned to the manufacturer for evaluation.
 
Additional Manufacturer Narrative
Returned for evaluation was one 16f valved peelable sheath.A visual review of the returned device noted that the hub was detached and in two pieces and the sheath was peeled.The device was forwarded to angiodynamics' vendor for analysis.The vendor performed an evaluation and discovered that the tube was not properly positioned during molding.The customers reported complaint description of the sheath fracturing is confirmed.It was noted the patient suffered no permanent harm or injury and is in stable condition.A review of the lot history records was performed for the reported packaging and component lots for any deviations related to the reported defect of the complaint.The review confirms that the lots met all material, assembly, and performance specifications the root cause, as stated by the vendor was "due to the tube not being properly positioned during molding, potentially related to an associate error in the positioning tube or core pins too smooth causing tube blow off during molding cycle." a review of the angiodynamics complaint system noted no adverse trends for this complaint type and product family.This type of complaint will continue to be monitored for trends.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DURAFLOW 2 CHRONIC HEMODIALYSIS CATHETER
Common Device Name
HEMODIALYSIS CATHETER
Manufacturer (Section D)
ANGIODYNAMICS, INC.
603 queensbury avenue
queensbury NY 12804
MDR Report Key5811754
Report Number1319211-2016-00101
Device Sequence Number10848089
Product Code MSD
UDI-Device IdentifierH787103012155
UDI-PublicH787103012155
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
K110936
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/22/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Expiration Date10/31/2018
Device Catalogue NumberH787103012155
Device Lot Number5004302
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/07/2016
Type of Report(Section G)FDA Requested
Initial Date Received by Manufacturer 06/22/2016
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date07/21/2016
Supplement Report FDA Received Date03/22/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-