MAUDE Adverse Event Report: MERGE HEALTHCARE MERGE HEMODYNAMICS; PROGRAMMABLE DIAGNOSTIC COMPUTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MERGE HEALTHCARE MERGE HEMODYNAMICS; PROGRAMMABLE DIAGNOSTIC COMPUTER Back to Search Results
Model Number MERGE HEMODYNAMICS 9.40.1
Device Problems Application Interface Becomes Non-Functional Or Program Exits Abnormally (1138); Failure to Transmit Record (1521); Device Issue (2379)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/23/2016
Event Type  malfunction  
Manufacturer Narrative
An internal investigation conducted by merge healthcare found that if the hemo monitor pc is powered up and there is no active network connection, the hemo monitor pc will report an appropriate error to the event log, but currently will not attempt any retries to establish the connection to the hemo client pc.No active network connection can occur if the cable between the hemo monitor pc and hemo client pc is disconnected, or if the hemo client pc has not been powered up.
 
Event Description
Merge hemodynamics monitors, measures, and records physiological data from a human patient undergoing a cardiac catheterization procedure.The system comprises the patient data module and the merge hemodynamics hemo monitor pc.The two units are connected via a serial interface.All vital parameters and evaluations are registered and calculated in the patient data module.This data is then transmitted to the merge hemodynamics hemo monitor pc via the serial interface.All data can be shown and monitored on the merge hemodynamics hemo monitor pc.On (b)(6) 2016, a customer reported to merge healthcare that hemo tabs were grayed out while a patient was present and the procedure was beginning.With merge hemo not presenting physiological data during treatment, there is a potential for a delay in care that results in harm to the patient.However, information provided by the customer indicates that the procedure was completed successfully once the hemo monitor was rebooted.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MERGE HEMODYNAMICS
Type of Device
PROGRAMMABLE DIAGNOSTIC COMPUTER
Manufacturer (Section D)
MERGE HEALTHCARE
900 walnut ridge drive
hartland WI 53029
Manufacturer (Section G)
MERGE HEALTHCARE
900 walnut ridge drive
hartland WI 53029
Manufacturer Contact
michael diedrick
900 walnut ridge drive
hartland, WI 53029
2629123570
MDR Report Key5818405
MDR Text Key50294590
Report Number2183926-2016-00658
Device Sequence Number1
Product Code DQK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K082421
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Remedial Action Recall
Type of Report Initial
Report Date 06/23/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMERGE HEMODYNAMICS 9.40.1
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 06/23/2016
Initial Date FDA Received07/24/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/24/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction Number2183926-02/15/2016-041-C