ATRIUM MEDICAL CORPORATION ATRIUM VEIN GRAFTTUNNELING SYSTEM; INSTRUMENTS, SURGICAL, CARDIOVASCULAR Back to Search Results
Model Number 26050
Health Effect - Clinical Code Death (1802)
Date of Event 04/25/2016
Type of Reportable Event Death
Additional Manufacturer Narrative
Associated file: 1219977-2016-00166.On completion, a follow up report will be submitted.
 
Event or Problem Description
The physician reported he was performing a femoral-popliteal bypass on a patient that resulted in the patient having excessive bleeding from the popliteal vein.The patient became hemodynamically unstable and was transferred to the intensive care unit.He expired on (b)(6) 2016 of multi-organ failure.
 
Additional Manufacturer Narrative
Engineering analysis: no lot numbers were provided so a review of manufacturing and qc records could not be performed.The territory sales manager performed a visual inspection of the tunneler kit; during this inspection he found no defects with the tunneler kit in question.The tips screwed on and off easily and had a smooth transition, no sharp edges or burrs were found.Photographs taken during this inspection have been reviewed and confirmed that there are no defects with the tunneler kit in question.Engineering summary: the engineering analysis could not determine the root cause of the incident.No defects were found during the inspection of the tunneler kit.All tunneler components are received with a certificate of compliance from the supplier.The tunneler kit instructions for use contain the following precautions: avoid damaging the threaded portions of the rods and tips through improper handling.The maquet tunneler requires disassembly, thorough cleaning, inspection and sterilization prior to each use.Do not soak the instruments in saline or bleach solutions, which may cause damage to the surfaces of the instruments.Do not use tunneler rod without compatible tip.Prior to use inspect instruments for the presence of foreign material including any residual tissue or debris, or any burrs or other surface anomalies that might injure the patient.Conclusion: based on the details of the event and the successful lot qualification test data, atrium can find no fault with the lot of tunneler kits in question.Clinical evaluation: the patient that requires a vascular graft has vascular tissue that may be hardened and calcified or very friable.Grafts are implanted surgically via an open incision and/or sub-cutaneous tunneling in certain applications.Surgeons undertaking such procedures are specialists in vascular disease and surgery and they have the skills necessary to be working in an operative suite and are supported by trained staff.In the event that an adverse situation occurs that is unexpected, an additional intervention may need to be performed.Any implantable device should be examined by the physician immediately prior to implant to confirm that the product is intact, that the threads of the component are undamaged and that the appropriate size is available.It is advised in the ifu of the product to contact atrium if these issues are noted.The ifu also advises that "avoid damaging the threaded portions of the rods and tips through improper handling.Tunneling can be one of the most traumatic parts of performing vascular surgery.Atrium designed its vascular tunneler system to minimize the trauma associated with tunneling.The atrium tunneler makes a tunnel that is the same size as graft being placed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATRIUM VEIN GRAFTTUNNELING SYSTEM
Common Device Name
INSTRUMENTS, SURGICAL, CARDIOVASCULAR
Manufacturer (Section D)
ATRIUM MEDICAL CORPORATION
5 wentworth drive
hudson NH 03051
MDR Report Key5825292
Report Number1219977-2016-00165
Device Sequence Number435414
Product Code DWS
UDI-Device Identifier00650862260502
UDI-Public00650862260502
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/05/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Physician
Device Model Number26050
Device Catalogue Number26050
Was Device Available for Evaluation? No
Type of Report(Section G)FDA Requested
Initial Date Received by Manufacturer 07/05/2016
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date07/27/2016
Supplement Report FDA Received Date08/10/2016
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Death;
Patient Age76 YR
-
-