• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ANIMAS CORPORATION ANIMAS VIBE INSULIN INFUSION PUMP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANIMAS CORPORATION ANIMAS VIBE INSULIN INFUSION PUMP Back to Search Results
Device Problem Poor Quality Image (1408)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
The pump has been returned to animas. Evaluation has not yet been completed. When evaluation is complete a supplemental report will be filed. No conclusion can be made at this time.
 
Event Description
On (b)(6) 2016, the reporter contacted animas, alleging a display (dim/fading/color spectrum) issue. This complaint is being reported because the issue may impact the user's ability to read some or all of the information on the screen which may result in over or under delivery. There was no indication that the product caused or contributed to an adverse event.
 
Manufacturer Narrative
Follow-up #1: date of submission 08/10/2016 - device evaluation: the device has been returned and evaluated by product analysis on 08/02/2016 with the following findings: during testing, the pump was powered on and the display screen appeared dim and discolored. The complaint was duplicated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameANIMAS VIBE
Type of DeviceINSULIN INFUSION PUMP
Manufacturer (Section D)
ANIMAS CORPORATION
200 lawrence dr
west chester PA 19380 3428
Manufacturer (Section G)
ANIMAS CORPORATION
200 lawrence dr
west chester PA 19380 3428
Manufacturer Contact
karin sargrad
200 lawrence dr
west chester, PA 19380-3428
4843561808
MDR Report Key5828304
MDR Text Key51820092
Report Number2531779-2016-17684
Device Sequence Number1
Product Code MDS
Combination Product (y/n)N
PMA/PMN Number
P130007
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign
Type of Report Initial,Followup
Report Date 07/07/2016
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received07/27/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/22/2016
Was the Report Sent to FDA? Yes
Device Age40 MO
Event Location No Information
Date Manufacturer Received07/07/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/12/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unkown

-
-