| Model Number |
N/A |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
Chemical Exposure (2570)
|
| Date of Event |
06/22/2016
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Blank fields on this form indicate the information is unknown or unavailable.This event is currently under investigation.
|
| |
|
Event or Problem Description
|
|
The devices are not waterproof.The radiologist ends up with the chimio on his hands therefore the patient has less treatment.Per the customer complaint form:"leak of product".The patient did not require any additional procedures due to this occurrence.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation- a review of the complaint history, device history record, quality control and visual inspection of the returned devices was conducted.No events related to the reported issue were found during a review of related manufacturing or quality records.Review of the associated device history record did not identify non-conformances that may be related to this reported incident.Approx 19 high pressure three-way stopcock of the reported lot number were returned for inspection; all were sealed and unused.The 19 stopcocks were visually examined and no defects were noted.Each stopcock was liquid tested and all passed.No leakage was observed.The device functioned as expected; there is no evidence to suggest the product was not manufactured to current specifications.The stopcocks utilized during the reported event were not returned for inspection; as a result the root cause is undetermined at this time.The proper personnel have been notified of this complaint, we will continue to monitor for similar issues.
|
| |
|