• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ANIMAS CORPORATION ANIMAS VIBE; INSULIN INFUSION PUMP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANIMAS CORPORATION ANIMAS VIBE; INSULIN INFUSION PUMP Back to Search Results
Device Problem Device Stops Intermittently (1599)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  No Answer Provided  
Manufacturer Narrative
The pump has not been returned to animas.If the device is returned, an evaluation shall be completed and a supplemental report will be filed.No conclusions can be made at this time.(b)(6).
 
Event Description
On (b)(6) 2016, the reporter contacted animas alleging an intermittent power issue with the pump.There is no additional information available for this complaint.There was no reported adverse event associated with this complaint.This complaint is reportable because the user may be unaware that the pump has lost power, leading to under delivery.
 
Manufacturer Narrative
Follow-up #1: date of submission 10/07/2016 device evaluation: the device has been returned and evaluated by product analysis on 09/20/2016 with the following findings: a review of the black box revealed unexpected power on reset (por) events on (b)(6) 2016.The battery compartment was observed to be cracked below the grip pad.There was no damage found to the returned battery cap; the battery cap secured tightly to the pump.The battery cap contact height and width measurements were within specification.During testing, the battery cap was fastened and then unscrewed ½ turn with no reboots occurring.The ¿ez-prime¿ steps were performed correctly with no power interruptions or any errors, alarms or warnings.The pump¿s cover was removed with no intermittent conditions found to the power circuit.Animas has conducted a review of the device history record for this pump and confirmed that it was operating within required specifications at the time of release.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANIMAS VIBE
Type of Device
INSULIN INFUSION PUMP
Manufacturer (Section D)
ANIMAS CORPORATION
200 lawrence dr
west chester PA 19380 3428
Manufacturer (Section G)
ANIMAS CORPORATION
200 lawrence dr
west chester PA 19380 3428
Manufacturer Contact
karin sargrad
200 lawrence dr
west chester, PA 19380-3428
4843561808
MDR Report Key5844244
MDR Text Key52439267
Report Number2531779-2016-18728
Device Sequence Number1
Product Code MDS
Combination Product (y/n)N
PMA/PMN Number
P130007
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign
Reporter Occupation Patient
Type of Report Initial,Followup
Report Date 07/14/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/03/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/12/2016
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? Yes
Device Age15 MO
Date Manufacturer Received07/14/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/10/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-