Device Problem
Material Discolored (1170)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The pump has not been returned to animas for evaluation.If the device is returned, an evaluation shall be completed and a supplemental report will be filed.No conclusions can be made at this time.(b)(6).
|
|
Event Description
|
On (b)(6) 2016, the reporter contacted animas, alleging a display (dim/fading/color spectrum) issue.There was no indication that the product caused or contributed to an adverse event.This complaint is reportable because the issue may impact the user's ability to read some or all of the information on the screen which may result in over or under delivery.
|
|
Manufacturer Narrative
|
Follow-up #1: date of submission 10/12/2016 device evaluation: the device has been returned and evaluated by product analysis on 09/21/2016 with the following findings: a dim display was observed with reddish text.Unrelated to the complaint, a cracked battery compartment was observed from the threads to the case seal left of the grip pad.Animas has conducted a review of the device history record for this pump and confirmed that it was operating within required specifications at the time of release.
|
|