No medical records or no medical images have been made available to the manufacturer.The device has been returned to the manufacturer for evaluation.As the lot number for the device was provided, a review of the device history records is currently being performed.The investigation of the reported event is currently underway.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
|
|
Manufacturing review: the device history records have been reviewed with special attention to the subassemblies, material review reports, raw material testing, manufacturing process, and quality control inspection.This lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.This is the only complaint that has been reported for this corporate lot number to date.Visual/microscopic inspection: the safety clip was missing upon receipt.The tuohy borst valve was loose and the two-way stop cock was open.The gray outer sheath was torn off approximately 55mm from the proximal end of the handle at the catheter reinforcement area.Stretching was noted around the break in the catheter.Exposed metal braiding was noted 236mm from the distal tip of the device.The stent graft was in place and not released.There was a gap between the atraumatic tip and the distal end of the outer catheter of 1 mm.Dried fluid was present between stent graft and outer catheter.Functional/performance evaluation: functional testing could not be performed, as the delivery system was returned with the outer sheath broken.Medical records review: medical records were not provided; therefore, a review could not be performed.Image/photo review: two digital photos were reviewed.The first image showed what appear to be the flair delivery system with a stent graft in place.A break was noted in the outer grey catheter of the delivery system in the photo.The second photo showed the product label.Based off the returned photos the investigation could be confirmed for failure to deploy due to a catheter breakage.Conclusion: the investigation was confirmed for deployment failure due to catheter breakage, as the stent graft was still in place in the delivery system, and the outer sheath was elongated and torn off at the catheter reinforcement area.Per the reported event details, the physician did not use a sheath.The ifu states under ¿materials required for device placement¿ that a 9f introducer sheath is required for device placement.Therefore, it is possible that procedural issues contributed to the reported event.However, the definitive root cause could not be determined.Labeling review: the current ifu (instructions for use) state: indication for use: flair endovascular stent graft is indicated for use in the treatment at the venous anastomosis of eptfe or other synthetic arteriovenous (av) access grafts.Warnings: the stent graft (implant) cannot be repositioned after total or partial deployment.Once partially or fully deployed, the flair endovascular stent graft cannot be retracted or remounted onto the delivery system.Device removal after deployment can only be done with a surgical approach.Precautions: the safety and effectiveness of this device have not been established when used around a tight bend in a looped av graft.It is recommended to access the av graft at the venous side of the av graft or at the apex.Careful attention should be paid to ensure the device is appropriately sized to the actual graft diameter, taking into account any change to the stated graft diameter that may have resulted from previous interventions.The appropriate length device(s) should be selected so that the stent graft extends beyond the stenosis into at least 10 mm of the non-diseased graft towards the arterial inflow and into the non-diseased vein approximately 10 mm beyond the stenosis.During deployment of the stent graft, the entire length of the delivery system should be kept as straight as possible in order to mitigate the potential for distal stent graft misplacement.After full stent graft deployment, wait a few seconds to allow for complete device expansion before removing the delivery system over the guidewire.Stent graft dislodgement has been reported during removal of the delivery system; therefore, careful attention should be paid during this portion of the procedure to prevent such occurrences.In case of placement of two stent grafts (overlap), use the same device diameter in both cases.If a flared device is used to overlap, do not deploy the flared end inside the first stent graft.Ensure a minimum 10 mm overlap of the devices.Potential complications and adverse events: stent graft specific events that could be associated with clinical complications include stent graft misplacement, stent graft migration, stent graft fracture, stent graft kinking, insufficient stent graft expansion and stent graft embolism.Delivery system specific events that could be associated with clinical complications include bond joint failures, detachment of parts, incompatibility with accessory devices, premature deployment, inaccurate deployment, failure to deploy, high deployment forces, delivery system kinking, no visibility under fluoroscopy, inability to track to target location, and blood leakage from delivery system (hemostasis).Stent graft size selection: special care must be taken to ensure that the appropriate size flair endovascular stent graft is selected.The stent graft body diameter should be approximately 1 mm larger than the synthetic av access graft diameter.(b)(4).The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
|