ARTHREX, INC. IBALANCE TKA BEARING; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
|
Back to Search Results |
|
| Catalog Number |
AR-513-BE08 |
| Health Effect - Clinical Code |
Pain (1994)
|
| Date of Event |
07/11/2016
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
No further patient information was provided at the time of this report or made available in response to arthrex's follow-up requests.This is the second of two submissions from the same patient event.The other is line 166377 (1220246-2016-00316).The device was requested/is expected for evaluation but has not yet been received.The contribution of the device to the reported event could not be determined as the device was not returned for evaluation therefore the complainant's event could not be verified.The cause of the event could not be determined from the information available and without device evaluation.Device history record review revealed nothing relevant to this event.If additional relevant information is received, a follow-up report will be submitted.Device being returned.Not yet received.
|
| |
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2015, patient underwent a left tka procedure.On (b)(6) 2016 surgeon performed a revision left tka due to patient pain.During the revision procedure the surgeon explanted the tibial tray ar-503-ttte (lot 1297366 )(cc92723-line 166377), and bearing implant ar-513-be08 (lot 113601337a)(cc92723-line 166378).To complete the procedure the surgeon implanted the following arthrex parts: tibial tray ar-513-t4 (lot 10038080), bearing implant ar-513-be13 (lot 113601414) and stem extension implant ar-513-1450 (lot 10036355).Patient was a female, (b)(6), dob (b)(6) 1955.During the revision procedure when the ar-613-1 universal handle lot 1391544 (cc92723-line 166654) was being used for the removal of the tibial punch head,.The punch head was able to be removed but the handle broke in the process which did not result in patient harm.Handle is being returned for evaluation.Patient medical records of (b)(6) 2016 noted that patient was reporting continued knee pain and swelling.Patient was injected and aspirated at that time.On (b)(6) 2016 records noted that injection/aspiration helped with symptoms for only a few days.Examination of the left knee demonstrated swelling.Palpation of the left knee demonstrated inferior pole patella tenderness, medial joint line and medial tibial plateau tenderness.Mild effusion of the left knee was noted.Records also indicated patient had pain with flexion and no pain with extension.Patient had a positive patellar grind test.Patient was scheduled for the (b)(6) 2016 revision.
|
| |
|
Additional Manufacturer Narrative
|
|
No further patient information was provided at the time of this report or made available in response to arthrex's follow-up requests.This is the second of two submissions from the same patient event.The other is line 166377 (1220246-2016-00316).The evaluation revealed that the returned implant has inconstant wear patterns between the condyles on the articulating surface.Wear pattern on one condyle is anterior and the other condyle is posterior.In addition, the c surface and the post have wear patterns.The damages observed on the locking tabs were most likely caused during the extraction of the implant.At the current time the cause of the event cannot be determined.If additional relevant information is received, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2015, patient underwent a left tka procedure.On (b)(6) 2016 surgeon performed a revision left tka due to patient pain.During the revision procedure the surgeon explanted the tibial tray ar-503-ttte (lot 1297366 )(cc92723-line 166377), and bearing implant ar-513-be08 (lot 113601337a)(cc92723-line 166378).To complete the procedure the surgeon implanted the following arthrex parts: tibial tray ar-513-t4 (lot 10038080), bearing implant ar-513-be13 (lot 113601414) and stem extension implant ar-513-1450 (lot 10036355).Patient was a female, (b)(6).During the revision procedure when the ar-613-1 universal handle lot 1391544 (cc92723-line 166654) was being used for the removal of the tibial punch head,.The punch head was able to be removed but the handle broke in the process which did not result in patient harm.Handle is being returned for evaluation.Patient medical records of (b)(6) 2016 noted that patient was reporting continued knee pain and swelling.Patient was injected and aspirated at that time.On (b)(6) 2016 records noted that injection/aspiration helped with symptoms for only a few days.Examination of the left knee demonstrated swelling.Palpation of the left knee demonstrated inferior pole patella tenderness, medial joint line and medial tibial plateau tenderness.Mild effusion of the left knee was noted.Records also indicated patient had pain with flexion and no pain with extension.Patient had a positive patellar grind test.
|
| |
|