ARTHREX, INC. BONE CUTTER,5.5MM X 13CM; BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL Back to Search Results
Catalog Number AR-8550BC
Health Effect - Clinical Code Device Embedded In Tissue or Plaque (3165)
Date of Event 07/18/2016
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Patient demographics (age at time of event, date of birth, gender, weight) were requested but not provided.No further patient information was provided at the time of this report or made available in response to follow-up communication.No additional adverse consequences have been reported from this event.This device is used for treatment.The device was requested for evaluation but was not returned, therefore the complainant's event could not be verified.The cause of the event could not be determined from the information available and without device evaluation.Device history record review revealed nothing relevant to this event.Complainant's event typically caused by user mechanical damage to device such as hitting the device with another device or excessive bending forces being applied during use.This is the first complaint of this type for this part/lot combination.The potential cause(s) of this event will be communicated to the event reporter.If additional relevant information is received, a follow-up report will be submitted.
 
Event or Problem Description
It was reported that the shaver caused metal debris.This was also visible under post-surgical mri.Shavings remained in the patient.Surgery performed was a sub-acrmoalia decompression.Surgery was successfully finished,using another identical shaver.
 
Additional Manufacturer Narrative
No further patient information was provided at the time of this report or made available in response to follow-up communication.No additional adverse consequences have been reported from this event.This device is used for treatment.Complaint confirmed.The evaluation revealed metal gouging on the id of outer tube and od of inner tube by distal end.Device history record review revealed nothing relevant to this event.Complainant's event typically caused by user mechanical damage to device such as hitting the device with another device or excessive bending forces being applied during use.This is the first complaint of this type for this part/lot combination.The potential cause(s) of this event will be communicated to the event reporter.If additional relevant information is received, a follow-up report will be submitted.
 
Event or Problem Description
It was reported that the shaver caused metal debris.This was also visible under post-surgical mri.Shavings remained in the patient.Surgery performed was a subacromial decompression.Surgery was successfully finished,using another identical shaver.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BONE CUTTER,5.5MM X 13CM
Common Device Name
BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
MDR Report Key5856341
Report Number1220246-2016-00333
Device Sequence Number16247417
Product Code GFA
UDI-Device Identifier00888867043930
UDI-Public00888867043930
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial,Followup
Report Date (Section B) 08/24/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date05/01/2021
Device Catalogue NumberAR-8550BC
Device Lot Number10044211
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/16/2016
Type of Report(Section G)FDA Requested
Initial Date Received by Manufacturer 07/20/2016
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date08/09/2016
Supplement Report FDA Received Date08/25/2016
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-