• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZOLL MEDICAL CORPORATION M SERIES; DEFIBRILLATOR/PACEMAKER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZOLL MEDICAL CORPORATION M SERIES; DEFIBRILLATOR/PACEMAKER Back to Search Results
Model Number MSERIES CCT
Device Problem Poor Quality Image (1408)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
The device was returned to zoll medical corporation.During evaluation of the device to determine root cause, the technician observed extensive damage.The observed damage is typically a result of the device being tampered with.Root cause could not be firmly established due to the observed damage.The device was recertified and returned to the customer.Analysis of reports of this type has not identified an increase in trend.
 
Event Description
Complainant alleged that during biomed testing, the device's display showed horizontal red and green lines and was not legible.Complainant indicated that there was no patient involvement in the reported malfunction.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
M SERIES
Type of Device
DEFIBRILLATOR/PACEMAKER
Manufacturer (Section D)
ZOLL MEDICAL CORPORATION
269 mill road
chelmsford MA 01824
Manufacturer Contact
269 mill road
chelmsford, MA 01824
9784219552
MDR Report Key5857958
MDR Text Key51746938
Report Number1220908-2016-01866
Device Sequence Number1
Product Code MKJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K011865
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Reporter Occupation Other
Type of Report Initial
Report Date 07/20/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/09/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberMSERIES CCT
Device Catalogue NumberM SERIES
Other Device ID Number00847946004811
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/22/2016
Is the Reporter a Health Professional? No
Date Manufacturer Received07/20/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/01/2005
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-