• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PENUMBRA, INC. PX SLIM DELIVERY MICROCATHETER; DQY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PENUMBRA, INC. PX SLIM DELIVERY MICROCATHETER; DQY Back to Search Results
Catalog Number PXSLIMSTR-A
Medical Device Problem Codes Kinked (1339); Mechanical Problem (1384); Physical Resistance (2578)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 07/11/2016
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Results: the px slim delivery microcatheter (px slim) was kinked approximately 5.0 cm from the distal tip.Conclusions: evaluation of the returned devices revealed the px slim was kinked in the distal shaft.This damage may have occurred due to forceful manipulation of the px slim during positioning in the patient anatomy.Further evaluation revealed the pc400 coil pusher assembly was kinked.This damage may have occurred due to forceful advancement of the pc400 coil against resistance.The kink in the px slim likely contributed to any resistance experienced during advancement of the pc400 coil.All penumbra coils and catheters are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.This report is associated with mfr report number: 3005168196-2016-01145.
 
Event or Problem Description
The patient was undergoing a coil embolization procedure in the cerebral artery using penumbra coil 400 coils (pc400 coils) and a px slim delivery microcatheter (px slim).During the procedure, while advancing a pc400 coil through a px slim, the physician experienced resistance and was unable to advance the pc400 coil out of the px slim.The pc400 coil was re-sheathed and re-introduced into the px slim.However, the physician was still unable to advance the pc400 coil out of the px slim and the pc400 coil pusher assembly became slightly bent.Therefore, the pc400 coil and px slim were removed.The physician also reported that there may have been a kink at the distal end of the px slim.The procedure was successfully completed using a new px slim and eight new pc400 coils.There was no report of an adverse effect to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PX SLIM DELIVERY MICROCATHETER
Common Device Name
DQY
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key5858190
Report Number3005168196-2016-01144
Device Sequence Number16235685
Product Code DQY
UDI-Device Identifier00814548016436
UDI-Public00814548016436
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
K100826
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 07/11/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Expiration Date12/31/2018
Device Catalogue NumberPXSLIMSTR-A
Device Lot NumberF66751
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/18/2016
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/11/2016
Initial Report FDA Received Date08/09/2016
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age39 YR
-
-