ANGIOMED GMBH & CO. MEDIZINTECHNIK KG FLUENCY PLUS ENDOVASCULAR STENT GRAFT Back to Search Results
Catalog Number FEL10060
Medical Device Problem Codes Fracture (1260); Activation, Positioning or Separation Problem (2906)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 07/13/2016
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device history records are being reviewed.The event is currently under investigation.The information provided represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant was unable to provide any patient details to date.
 
Event or Problem Description
It was reported that during deployment of the endovascular stent graft, the outer sheath of the delivery system fractured; however, the stent graft could be deployed successfully.There was no reported patient injury.
 
Additional Manufacturer Narrative
The lot history records have been reviewed with special attention to the manufacturing and inspection of this product and the product was found to have met all specifications prior to shipment.There is no indication for a manufacturing-related failure.On the basis of the evaluation of the returned device, the reported event could be confirmed.The outer sheath was found to be elongated and fractured indicating the presence of high deployment force during the attempt to deploy the stent graft.The stent graft was not returned as it was successfully implanted.Potential contributing factors to the reported event have been considered.Previous investigations of similar complaints have been reviewed.The reported event may be associated with a difficult vessel anatomy leading to increased friction during deployment and subsequent sheath fracture.Insufficient flushing of the device may be another contributing factor to the reported event.On the basis of the information available, a definite root cause for the reported event could not be determined.The ifu states that the device must be flushed with sterile saline.Also the ifu states: "the safety and effectiveness of the device when placed across a tight bend (.) has not been evaluated.Prior to stent graft deployment, ensure that the proximal (inflow) stent graft end is positioned in a straight section of the lumen to reduce the risk of higher deployment forces and possible endovascular system failure." and "do not kink the delivery catheter or use excessive force during delivery to the target lesion." furthermore, the ifu states: "if unusual resistance or high deployment force is encountered during stent graft deployment, abort the procedure, remove the delivery system and use an alternative device." evaluation was completed.
 
Event or Problem Description
It was reported that during deployment of the endovascular stent graft, the outer sheath of the delivery system fractured; however, the stent graft could be deployed successfully.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLUENCY PLUS ENDOVASCULAR STENT GRAFT
Common Device Name
ENDOVASCULAR STENT GRAFT
Manufacturer (Section D)
ANGIOMED GMBH & CO. MEDIZINTECHNIK KG
wachhausstrasse 6
karlsruhe 76227
GM  76227
MDR Report Key5864664
Report Number9681442-2016-00221
Device Sequence Number1230521
Product Code PFV
UDI-Device Identifier04049519008240
UDI-Public(01)04049519008240(17)180627(10)ANZF3167
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
P130029
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/15/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/27/2018
Device Catalogue NumberFEL10060
Device Lot NumberANZF3167
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/05/2016
Type of Report(Section G)FDA Requested
Initial Date Received by Manufacturer 07/15/2016
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date08/10/2016
Supplement Report FDA Received Date09/28/2016
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-