• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK INC ADVANCE 14 LP LOW PROFILE BALLOON CATHETER; LIT CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC ADVANCE 14 LP LOW PROFILE BALLOON CATHETER; LIT CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL Back to Search Results
Model Number N/A
Medical Device Problem Code Torn Material (3024)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).This event is currently under investigation.
 
Event or Problem Description
During a tibial angioplasty procedure, the wire got stuck due to the aortic arch and it ripped the exchange port of the delivery catheter.A new advance 14 lp low profile balloon catheter was opened and used to complete the procedure.A section of the device did not remain inside the patient's body.According to the initial reporter, the patient did not require any additional procedures nor experience any adverse effects due to this occurrence.
 
Additional Manufacturer Narrative
(b)(4).Investigation - evaluation: a review of the complaint history, drawings, device history record, documentation, manufacturing instructions and quality control was conducted during the investigation.The complaint device was not returned therefore, no physical examinations could be performed; however, a document based investigation evaluation was performed.There is no evidence to suggest the product was not made to specifications.Review of device history record shows no nonconforming events which could contribute to this failure mode.It should be noted there were no other reported complaints for this lot number.Tensile testing is performed during the manufacturing process.It is likely that excess force contributed to this failure mode, however, without objective evidence, the root cause of this failure mode is unknown.Based on the information provided, no product returned and the results of our investigation, a definitive root cause could not be determined.We will continue to monitor for similar complaints.Per the risk assessment no further action is required.
 
Event or Problem Description
During a tibial angioplasty procedure, the wire got stuck due to the aortic arch and it ripped the exchange port of the delivery catheter.A new advance 14 lp low profile balloon catheter was opened and used to complete the procedure.A section of the device did not remain inside the patient's body.According to the initial reporter, the patient did not require any additional procedures nor experience any adverse effects due to this occurrence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANCE 14 LP LOW PROFILE BALLOON CATHETER
Common Device Name
LIT CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
MDR Report Key5868147
Report Number1820334-2016-00649
Device Sequence Number18940222
Product Code LIT
UDI-Device Identifier10827002503356
UDI-Public(01)10827002503356(17)180312(10)5715154
Combination Product (Y/N)N
Initial Reporter StateCT
Initial Reporter CountryUS
PMA/510(K) Number
K090822
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/14/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue NumberPTAX4-14-170-4-4
Was Device Available for Evaluation? No
Type of Report(Section G)FDA Requested
Initial Date Received by Manufacturer 07/14/2016
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date08/11/2016
Supplement Report FDA Received Date11/09/2016
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-