COOK ENDOSCOPY FUSION® PRE-LOADED WITH ACROBAT WIRE GUIDE; KNS, UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES) Back to Search Results
Catalog Number FS-OMNI-ACRO-35-205
Medical Device Problem Code Peeled/Delaminated (1454)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 07/19/2016
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Concomitant medical products: fujinon endoscope, model #5d102a515, 4.2 mm working channel.Investigation evaluation: a product evaluation was not performed in response to this report because the product said to be involved was not provided to cook for evaluation.The report could not be confirmed.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: we could not conduct a complete investigation because the product said to be involved was not returned for evaluation.A definitive cause for the reported observation could not be determined.The instructions for use advise the user that "for best results, wire guide should be kept wet." the instructions for use describe the appropriate flushing techniques for use of this coated wire guide.These techniques include flushing the endoscope accessory channel and/or lumen of accessory device with sterile water before wire guide insertion.If these flushing techniques are not followed or inadequate flushing occurs, this can contribute to wire guide coating damage.Prior to distribution, all fusion pre-loaded with acrobat wire guides are subjected to a visual inspection and functional test to ensure device integrity.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
 
Event or Problem Description
During an endoscopic procedure, the physician used a cook fusion pre-loaded with acrobat wire guide.The wire guide stripped from approximately 25 cm from distal tip.The physician ran a lithotripsy device down the working channel alongside the wire guide [lithotripsy device was not advanced over wire guide].The case was finished and no harm caused to patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FUSION® PRE-LOADED WITH ACROBAT WIRE GUIDE
Common Device Name
KNS, UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES)
Manufacturer (Section D)
COOK ENDOSCOPY
4900 bethania station rd
winston-salem NC 27105
MDR Report Key5870075
Report Number1037905-2016-00293
Device Sequence Number16293017
Product Code KNS
UDI-Device Identifier10827002346014
UDI-Public(01)10827002346014(17)190331(10)W3697006
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K052051
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Nurse
Type of Report Initial
Report Date (Section B) 07/19/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberFS-OMNI-ACRO-35-205
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Distributor Facility Aware Date07/19/2016
Device Age4 MO
Event Location Hospital
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/19/2016
Initial Report FDA Received Date08/11/2016
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-