• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL, L.P. DELTA HIP; HEAD, FEMORAL, CERAMIC, BILOX DELTA, 36MM NEUTRAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL, L.P. DELTA HIP; HEAD, FEMORAL, CERAMIC, BILOX DELTA, 36MM NEUTRAL Back to Search Results
Catalog Number 400-03-362
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 07/15/2016
Type of Reportable Event Serious Injury
Event or Problem Description
Revision surgery - due to an infection.
 
Additional Manufacturer Narrative
Corrected data: brand name, common device name, model/lot #, and concomitant medical products.The concomitant and suspect medical device were reported incorrectly on the initial medical device report, they were switched.Manufacturer narrative: the reason for this revision surgery was an infection.The previous surgery and the revision detailed in this investigation occurred 24 days apart.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the possible infection.The healthcare professional indicated there was a significant adverse event to the patient.There was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the implant device history records (dhrs) and product complaint report (pcr) database records show that the reported component used in the previous surgery met design and manufacturing requirements.There were no non-conforming material reports (ncmrs) associated with the product that may have contributed to an infection.The device was verified to have gone through an acceptable sterilization process and was within it's expiration date at the time of use during the previous surgery.There were no findings during this investigation that indicate that the reported device was the root cause or had a direct connection with the patient's infection.Due to the short time between the original surgery and the revision, it is possible that the infection was acquired in the hospital (nosocomial).It is also possible that the patient was not compliant with post surgical instructions.There are multiple factors that may contribute to an infection that are outside of the control of djo surgical.There are no indications of a product or process issue affecting implant safety or effectiveness.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DELTA HIP
Common Device Name
HEAD, FEMORAL, CERAMIC, BILOX DELTA, 36MM NEUTRAL
Manufacturer (Section D)
ENCORE MEDICAL, L.P.
9800 metric blvd.
austin TX 78758
MDR Report Key5873499
Report Number1644408-2016-00561
Device Sequence Number16279605
Product Code LZO
UDI-Device Identifier00888912100175
UDI-Public(01)00888912100175
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K082844
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2016
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/15/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/27/2021
Device Catalogue Number400-03-362
Device Lot Number866B1197
Was Device Available for Evaluation? No
Type of Report(Section G)FDA Requested
Initial Date Received by Manufacturer 07/15/2016
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date08/12/2016
Supplement Report FDA Received Date09/22/2016
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Required Intervention;
Patient Age71 YR
-
-