Model Number ETH2B5LT
Device Problem Device Contamination with Chemical or Other Material (2944)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/13/2016
Event Type  malfunction  
Manufacturer Narrative
The device has not yet been returned to the manufacturer at the time of this report.A supplemental form will be sent once the evaluation is completed if the device is returned.The device history report was reviewed and no discrepancies were found.
 
Event Description
It was reported that during a procedure, debris was found in the device.It is reported that the patient was being monitored for adverse reaction, but at the time of this report, no patient harm or consequences were noted.
 
Manufacturer Narrative
The account supplied photos showing the debris inside the obturator, along with the debris missing from the camera view in the field.When the device was returned, the rear port was uncovered and no debris was found to be present.The debris in the photos supplied by the account is reddish/brown which is inconsistent with organic debris that has been exposed to eo sterilization.The photos supplied by the account confirm that debris was present during use in the field, though no analysis of the debris could be completed due to the debris being missing when returned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
LAPAROSCOPE, GENERAL & PLASTIC SURGERY, REPROCESSED
Manufacturer (Section D)
STERILMED, INC.
11400 73rd ave n
maple grove MN
Manufacturer Contact
jason anderson
5010 cheshire parway
suite 2
plymouth, MN 55446
7634888348
MDR Report Key5876396
MDR Text Key52111810
Report Number2134070-2016-00057
Device Sequence Number1
Product Code NLM
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K043592
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 09/05/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/22/2017
Device Model NumberETH2B5LT
Device Catalogue NumberETH2B5LT
Device Lot Number1918455
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/05/2016
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/19/2016
Initial Date FDA Received08/15/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received09/07/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/22/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-