Model Number MODEL 250
Device Problem Bent (1059)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/21/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
The tablet was reported broken at the usb port, which was visualized.The prongs on the usb port of the tablet were bent to the point that the serial cable had to held in an inconvenient position in order for it to work.No issues were seen with the physician's serial cable or wand.Additional information was received that the issue was first observed when a company representative visited the office on (b)(6) 2016.The handling of the device was not suspected to have caused the issue.The wand and serial cable was tested with the company representative's programming system to determine that there was no issue with their function.The tablet has not been received to date.
 
Event Description
The tablet was received on 8/15/2016.An analysis was performed on the returned tablet and the reported allegation of mechanical problem with usb cable to tablet connection was verified.During the analysis it was identified that the usb port was damaged.As a result, the tablet was unable to establish communication.No further anomalies were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGRAMMING SOFTWARE
Type of Device
PROGRAMMING COMPUTER
Manufacturer (Section D)
CYBERONICS, INC.
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
CYBERONICS, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key5877066
MDR Text Key52382519
Report Number1644487-2016-01834
Device Sequence Number1
Product Code LYJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 07/21/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/15/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 250
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/15/2016
Is the Reporter a Health Professional? No
Event Location Other
Date Manufacturer Received08/15/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/08/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-