| Catalog Number |
AR-1934BCF-24 |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
07/29/2016
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
No further patient information was provided at the time of this report or made available in response to follow-up communication.No additional adverse consequences have been reported from this event.This device is used for treatment.The contribution of the device to the reported event could not be determined as the device was not returned for evaluation therefore the complainant's event could not be verified.The cause of the event could not be determined from the information available and without device evaluation.Device history record review revealed nothing relevant to this event.This type of event is typically caused by not inserting the implant coaxial to the bone tunnel, prying/leveraging the driver while the implant is still loaded and/or improper bone preparation prior to insertion.This is the first complaint of this type for this part/lot combination.The potential cause(s) of this event will be communicated to the event reporter.If additional relevant information is received, a follow-up report will be submitted.Device remains in patient.
|
| |
|
Event or Problem Description
|
|
It was reported that during an orif right elbow coronoid fracture repair, the bio-composite suturetak suture anchor broke on insertion.The surgeon attempted multiple times to remove the anchor but was not able to do so.No patient injury reported.
|
| |
|