• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: APOLLO ENDOSURGERY, INC. LAP-BAND AP LARGE WITH ACCESS PORT II; ADJUSTABLE GASTRIC BAND

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

APOLLO ENDOSURGERY, INC. LAP-BAND AP LARGE WITH ACCESS PORT II; ADJUSTABLE GASTRIC BAND Back to Search Results
Model Number B-2265
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code Patient Problem/Medical Problem (2688)
Date of Event 06/20/2016
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Taper ii medwatch sent to the fda on 08/16/2016.The reporter of the event was asked to return the product for analysis.To date, apollo has not received the device.Based on the serial number provided, the connector type is assumed to be a taper ii.If returned, visual examination may confirm or determine another taper type associated with this event.Device labeling addresses the event of leaking port as follows: precautions: care must be taken during band adjustment to avoid puncturing the tubing that connects the access port and band, as this will cause leakage and deflation of the inflatable section.Failure to create a stable, smooth path for the access port tubing, without sharp turns or bends, can result in tubing breaks and leakage.In order to avoid incorrect placement, the port should be placed lateral to the trocar opening.A pocket must be created for the port so that it is placed far enough from the trocar path to avoid abrupt kinking of the tubing.The tubing path should point in the direction of the access port connector so that the tubing will form a straight line with a gentle arching transition into the abdomen.Adverse events: it is important to discuss all possible complications and adverse events with your patient.Complications which may result from the use of this product include the risks associated with the medications and methods utilized in the surgical procedure, the risks associated with any surgical procedure and the patient's degree of intolerance to any foreign object implanted in the body.Unplanned deflation of the band may occur due to leakage from the band, the port or the connecting tubing.Warnings: patients should be advised that the lap-band ap® system is a long-term implant.Explant (removal)and replacement surgery may be indicated at any time.Medical management of adverse reactions may include explantation.Revision surgery for explantation and replacement may also be indicated to achieve patient satisfaction.
 
Event or Problem Description
Reported as: a patient with the lap-band system was reported to have a "leaking port." the device was explanted.
 
Additional Manufacturer Narrative
Further review of this record noted this event had been reported on mdr 3006722112-2016-00282.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LAP-BAND AP LARGE WITH ACCESS PORT II
Common Device Name
ADJUSTABLE GASTRIC BAND
Manufacturer (Section D)
APOLLO ENDOSURGERY, INC.
1120 s capitol of texas hwy
bldg 1, ste 300
austin TX 78746
MDR Report Key5878658
Report Number3006722112-2016-00238
Device Sequence Number842356
Product Code LTI
UDI-Device Identifier10811955020237
UDI-Public10811955020237
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
P000008
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2014
Device Explanted Year2016
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/19/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/20/2016
Device Model NumberB-2265
Device Catalogue NumberB-2265
Device Lot Number2540388
Was Device Available for Evaluation? No
Type of Report(Section G)FDA Requested
Initial Date Received by Manufacturer 07/19/2016
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date08/16/2016
Supplement Report FDA Received Date03/17/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age47 YR
Patient Weight94
-
-