|
It was reported that the physician has been having intermittent programming issues with his programming system on (b)(6) 2016.The nurse didn't know what error messages were observed.She called to report the issues after the event and didn't have the programmer with her, so no troubleshooting could be performed on the phone.Additional information was received from a company representative that the serial cable on the tablet was replaced on (b)(6) 2016 and the programming system is now working fine.Per the company representative, the wand did have a good battery and the power lights stayed on for 25 seconds.The company representative tried removing the old white serial cable and reconnecting it after waiting for 15 seconds and attempted interrogation but it did not work.Therefore, the old white serial cable was replaced with a new black serial cable and the programming system then worked.No patients were affected.The serial cable does not require return, as the failure mode is understood to be a failure of the serial cable associated with a disconnected wire connection.
|