MAUDE Adverse Event Report: DEXCOM, INC. DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEXCOM, INC. DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM Back to Search Results
Model Number MT22430-PNK
Medical Device Problem Code No Audible Alarm (1019)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 07/25/2016
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
Dexcom was made aware on (b)(6) 2016, to claim no audio output on (b)(6) 2016.There was no report of any injury or medical intervention.No additional event or patient information is available.The complaint device was returned for evaluation.The device was visually inspected and no defects were found.A "try it" manual test was performed and speaker did not sound.The reported event of no audio output was confirmed.The root cause was determined to be a defective speaker.
 
Additional Manufacturer Narrative
(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM
Common Device Name
CONTINUOUS GLUCOSE MONITOR
Manufacturer (Section D)
DEXCOM, INC.
6340 sequence drive
san diego CA 92121
MDR Report Key5887560
Report Number3004753838-2016-49387
Device Sequence Number13701837
Product Code MDS
UDI-Device Identifier10386270000108
UDI-Public(01)10386270000108(241)MT22430-PNK(10)5195973(17)NI
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P120005
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Remedial Action Recall
Type of Report Initial,Followup
Report Date (Section B) 07/25/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMT22430-PNK
Device Catalogue NumberSTK-KD-PNK
Device Lot Number5195973
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/05/2016
Type of Report(Section G)FDA Requested
Initial Date Received by Manufacturer 07/25/2016
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date08/18/2016
Supplement Report FDA Received Date09/16/2016
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
3004753838-02/29/16-001C
Patient Sequence Number1
Patient Age13 YR
Patient Weight50
-
-