• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WILLIAM COOK EUROPE COOK CELECT NAVALIGN JUGULAR & FEMORAL VENA CAVA FILTERS; DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WILLIAM COOK EUROPE COOK CELECT NAVALIGN JUGULAR & FEMORAL VENA CAVA FILTERS; DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR Back to Search Results
Catalog Number IGTCFS-65-2-UNI-CELECT
Device Problem Positioning Failure (1158)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 03/01/2012
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Catalog # igtcfs-65-2-uni-celect.Similar to device under 510(k) k090140.Summary of investigational findings: the investigation is based on event description(s) and provided imaging.In imaging review of event, the filter configuration post-deployment indicates that the filter was deployed in a narrow ivc.No indications of perforation of ivc with the filter hook, since images show no obvious signs of extravasation.Images also indicate that the filter was pulled back into the sheath since the secondary filter legs are completely damaged and curled up around the filter hook.There is found no evidence to suggest that this device was not manufactured according to specifications.Cook medical will continue to monitor for similar events.
 
Event Description
Description of event according to complainant: filter was advanced but it perforated the ivc wall.Patient outcome: unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COOK CELECT NAVALIGN JUGULAR & FEMORAL VENA CAVA FILTERS
Type of Device
DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR
Manufacturer (Section D)
WILLIAM COOK EUROPE
4632 bjaeverskov
DA 
Manufacturer Contact
marianne hoey
sandet 6
bjaeverskov 4632
DA   4632
56868686
MDR Report Key5902790
MDR Text Key52947745
Report Number3002808486-2016-00983
Device Sequence Number1
Product Code DTK
UDI-Device Identifier10827002529219
UDI-Public(01)10827002529219(17)141027(10)E2822932
Combination Product (y/n)N
Reporter Country CodeGB
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 03/06/2012
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/25/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberIGTCFS-65-2-UNI-CELECT
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/09/2012
Date Device Manufactured10/01/2011
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-