Catalog Number 9731780
Device Problems Bent (1059); Material Deformation (2976)
Patient Problem No Patient Involvement (2645)
Event Date 07/29/2016
Event Type  malfunction  
Manufacturer Narrative
No patient information provided as no patient was involved in this concern.Return requested.Replacement open spine clamp shipped to site 08/10/2016.Medtronic investigation of returned suspect open spine clamp found slight damage of the threads that made it a little difficult at different spots, however, the clamp was still fully functional.Reported issue could not be replicated.No fault found.
 
Event Description
A medtronic representative reported a site's open spine clamp had a stripped screw and was bent.This was noted prior to the start of a procedure.No further details regarding the damage, or how it occurred, were provided.There was no patient present when this issue was identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPEN SPINE CLAMP, TITANIUM
Type of Device
NEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
judith ericson
826 coal creek circle
louisville, CO 80027-9710
7208902187
MDR Report Key5903407
MDR Text Key52969754
Report Number1723170-2016-03099
Device Sequence Number1
Product Code HAW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K990214
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial
Report Date 08/25/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number9731780
Device Lot Number130729
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/18/2016
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 07/29/2016
Initial Date FDA Received08/25/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/29/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-