CYBERONICS - HOUSTON PULSE GEN MODEL 106; GENERATOR Back to Search Results
Model Number 106
Device Problem High impedance (1291)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/01/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that high impedance was registered on the patient's vns system.The site performed a chest x-rays and reported to not show a visual anomaly with the lead.The device was disabled at that time and the patient was referred for surgery.The generator and lead dhrs were reviewed and found that all specifications met prior to distribution.The patient underwent surgery on (b)(6) 2016.At the start of the case, system diagnostics were performed that still showed high impedance.During the case, the test resistor was inserted into the generator, and a generator diagnostic showed expected values that were within normal limits.The lead was re-inserted and the impedance was measured to be within normal limits, 3718 ohms on one test and 3659 ohms on the second.The surgery was completed without replacing any products.The x-ray images taken prior to the case were reviewed by a company field representative, and the lead could not be visualized to be inserted past the generator connector block in the images.No additional pertinent information has been received to date.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PULSE GEN MODEL 106
Type of Device
GENERATOR
Manufacturer (Section D)
CYBERONICS - HOUSTON
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
CYBERONICS - HOUSTON
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key5906415
MDR Text Key54057430
Report Number1644487-2016-01927
Device Sequence Number1
Product Code LYJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 08/03/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/26/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date10/20/2017
Device Model Number106
Device Lot Number4593
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Event Location Other
Date Manufacturer Received08/03/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/28/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age17 YR
-
-