It was reported that high impedance was registered on the patient's vns system.The site performed a chest x-rays and reported to not show a visual anomaly with the lead.The device was disabled at that time and the patient was referred for surgery.The generator and lead dhrs were reviewed and found that all specifications met prior to distribution.The patient underwent surgery on (b)(6) 2016.At the start of the case, system diagnostics were performed that still showed high impedance.During the case, the test resistor was inserted into the generator, and a generator diagnostic showed expected values that were within normal limits.The lead was re-inserted and the impedance was measured to be within normal limits, 3718 ohms on one test and 3659 ohms on the second.The surgery was completed without replacing any products.The x-ray images taken prior to the case were reviewed by a company field representative, and the lead could not be visualized to be inserted past the generator connector block in the images.No additional pertinent information has been received to date.
|