• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL, L.P. FMP COCR (COBALT CHROMIUM) HEADS; HEAD, FEMORAL, COCR, 40MM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL, L.P. FMP COCR (COBALT CHROMIUM) HEADS; HEAD, FEMORAL, COCR, 40MM Back to Search Results
Catalog Number 497-40-000
Health Effect - Clinical Code Joint Dislocation (2374)
Date of Event 04/14/2016
Type of Reportable Event Serious Injury
Event or Problem Description
Revision surgery - due to chronic dislocation.
 
Additional Manufacturer Narrative
The reason for this revision surgery was the patient had chronic dislocation.The in-vivo length of service for the patient's implant was one day.There is no information in this complaint about any patient injuries, activities, or accidents that may have contributed to the need for this revision surgery.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the device history records (dhrs) revealed no discrepancies or issues with the manufacturing history of this part.All critical dimensions and specifications in effect at the time the part was manufactured were met.The product complaint report history was reviewed and no trends or on-going issues were deemed as present or in need of review.This event is deemed to be non-product related.The root cause for the dislocations was not reported.The scope of this investigation is limited without having the parts available to djo surgical for evaluation.Other conditions, root cause, relating to this event could not be determined with confidence.Inventory containment is not required as there are no indications of a product or process issue affecting implant safety or effectiveness.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FMP COCR (COBALT CHROMIUM) HEADS
Common Device Name
HEAD, FEMORAL, COCR, 40MM
Manufacturer (Section D)
ENCORE MEDICAL, L.P.
9800 metric blvd.
austin TX 78758
MDR Report Key5908633
Report Number1644408-2016-00675
Device Sequence Number952248
Product Code JDI
UDI-Device Identifier00888912024068
UDI-Public(01)00888912024068
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K932481
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2016
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/14/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date02/28/2021
Device Catalogue Number497-40-000
Device Lot Number641C1056
Was Device Available for Evaluation? No
Type of Report(Section G)FDA Requested
Initial Date Received by Manufacturer 04/14/2016
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date08/26/2016
Supplement Report FDA Received Date10/13/2016
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age59 YR
-
-