• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ACE SURGICAL SUPPLY COMPANY INTERNAL HEX IMPLANT; DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ACE SURGICAL SUPPLY COMPANY INTERNAL HEX IMPLANT; DENTAL IMPLANT Back to Search Results
Model Number 00614950007196 00614950007196 00614950007196 00614950007196 214711
Medical Device Problem Code Osseointegration Problem (3003)
Health Effect - Clinical Codes Failure of Implant (1924); Unspecified Infection (1930)
Date of Event 07/19/2016
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
This implant loss suggests that the source of the problem was likely to stem from procedural errors.Correct use and procedural instructions are described in this product's instructions for use.
 
Event or Problem Description
Infection was reported.Time of loss in relation to the implantation was before prosthetic restoration.Primary stability and osseointegration was not achieved.Bone quality at the time of implant failure was type iii.Implant did not achieve integration.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTERNAL HEX IMPLANT
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
ACE SURGICAL SUPPLY COMPANY
1034 pearl street
brockton MA 02301
MDR Report Key5921059
Report Number1287163-2016-00269
Device Sequence Number11011694
Product Code DZE
UDI-Device Identifier00614950007196
UDI-Public00614950007196
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K103790
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2016
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Dentist
Type of Report Initial
Report Date (Section B) 08/12/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Dentist
Device Model Number00614950007196 00614950007196 00614950007196 00614950007196 214711
Device Catalogue Number214711
Device Lot Number14080004
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/12/2016
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/12/2016
Initial Report FDA Received Date09/01/2016
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age75 YR
-
-