MAUDE Adverse Event Report: BOSTON SCIENTIFIC - MAPLE GROVE AGENT; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - MAPLE GROVE AGENT; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS Back to Search Results
Model Number H74939222352010
Device Problems Device Damaged Prior to Use (2284); Difficult to Open or Remove Packaging Material (2922)
Patient Problem No Patient Involvement (2645)
Event Date 06/07/2016
Event Type  malfunction  
Manufacturer Narrative
Device evaluated by manufacturer: returned product consisted of an agent dcb.The balloon was tightly folded and was not damaged.The tip/shaft bond was separated and was not returned for analysis.There were numerous kinks throughout the hypotube.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The most probable root cause is considered handling damage as the event occurred prior to patient contact.(b)(4).
 
Event Description
Reportable based on the product analysis completed on august 9th, 2016.During unpacking of the 3.50 mm x 20.00 mm agent drug coated balloon (dcb) catheter, the balloon seemed to be stuck to the plastic protector so when they removed the protector, the balloon was damaged.Different device was used to complete the case.However, the returned product analysis revealed that the distal tip was detached.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AGENT
Type of Device
CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC - MAPLE GROVE
one scimed place
maple grove MN 55311
Manufacturer Contact
linda leimer
one scimed place
maple grove, MN 55311
7634941700
MDR Report Key5921530
MDR Text Key53717307
Report Number2134265-2016-07980
Device Sequence Number1
Product Code LOX
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
SIMILAR
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 08/09/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/31/2017
Device Model NumberH74939222352010
Device Catalogue Number39222-35201
Device Lot Number6727H15
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/14/2016
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/09/2016
Initial Date FDA Received09/01/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/23/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-